Stent (Diameter to Length)
(From Diagnostic and Interventional Catheterization in Congenital Heart Disease. 2nd ed. by James E. Lock et al).
3/21/2015
Diabolo Stent
Diabolo stent placement technique (Fontan Fenestration)
O.Stumper et al. Heart 2003;89:1227-30.
Technique: 3-4 mm diameter loop is created using temporary epicardial pacing wires as follows: The needle ends are removed. 5 cm of the wire tips are bared of its insulation. The 2 bare wire ends are tied in a loop. This provides ~90 cm long wire. A secure double-knot over a 10-14 French dilator.
The loop is paced over a 15-25 mm Valvuloplasty balloon (?low pressure). J&J P308 or P188/JoMed 17 mm stents are options. The stent is gently dilated using 10-14 French dilator and then, mounted on the balloon - keeping the knotted-loop over the center of balloon.
Deliver using 11-14 Fr Mullin sheath.