5/16/2020

Stents - Maximum Expandable Diameters


(From Congenital Heart Disease 2015;10:51-63)

8/23/2019

12/16/2018

Stent - Over-expansion tests

From Bratincsak, A. et al. Congenital Heart Dis 2015;10:51-6.


8/04/2018

Stent (Diameter to Length)

(From Diagnostic and Interventional Catheterization in Congenital Heart Disease. 2nd ed. by James E. Lock et al).

7/19/2017

Stents Comparison: Radial Strength vs. Comformability

(Image from Boston Scientific Express LD coronary stent brochure)


3/21/2015

Diabolo Stent

Diabolo stent placement technique (Fontan Fenestration)
O.Stumper et al. Heart 2003;89:1227-30.

Technique: 3-4 mm diameter loop is created using temporary epicardial pacing wires as follows: The needle ends are removed. 5 cm of the wire tips are bared of its insulation. The 2 bare wire ends are tied in a loop. This provides ~90 cm long wire. A secure double-knot over a 10-14 French dilator.
The loop is paced over a 15-25 mm Valvuloplasty balloon (?low pressure). J&J P308 or P188/JoMed 17 mm stents are options. The stent is gently dilated using 10-14 French dilator and then, mounted on the balloon - keeping the knotted-loop over the center of balloon.
Deliver using 11-14 Fr Mullin sheath.
 



6/16/2011

Cath: Stents - Intrastent LD

Intrastent LD (large diameter) - With EverCross balloons (Manufacturer: eV3)
3 stents:
1) Intrastent DoubleStrut LD Biliary Stent (Double struts)
2) Intrastent Mega LD Biliary Stent (5 cell design)