Infant, s/p Repair of AVSD (Postop. day 8) Postoperative Heart Block Temporary pacemaker setting: DDD @ 90 bpm. What is the mistake made by the nurse? (or the resident, or the fellow...attending would not have done it!!)
2 comments:
Anonymous
said...
Pacing in inappropriate. Appears to be a spike followed by QRS, then spike followed by p wave.
2 comments:
Pacing in inappropriate. Appears to be a spike followed by QRS, then spike followed by p wave.
That's correct. The atrial and ventricular leads were plugged in the wrong outlet in the temporary pacemaker!